7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 219
mmmmm
Appendix C: Troubleshooting
Noise Problems
Incorrect readings can be the result of noise problems, either acoustic or electrical, in the
application.
The noise present at the input to the ultrasonic receiver can be determined by viewing
parameter P807. The display reads ##:##, where the first number is the average noise and
the second is the peak noise. In general, the most useful value is the average noise.
With no transducer attached the noise is under 5 dB. This is often called the noise floor. If
the value with a transducer attached is greater than 5 dB, then signal processing
problems can occur. High noise decreases the maximum distance that can be measured.
The exact relationship between noise and maximum distance is dependent on the
transducer type and the material being measured. Any average noise level greater than
20 dB is probably cause for concern unless the distance is much shorter than the
maximum for the transducer.
Reading fluctuates
while material level is
still (or vice versa).
Incorrect measurement stabi-
lization.
Alter Maximum Process Speed
(P003) or damping (P704)
accordingly. See Maximum Pro-
cess Speed.
Reading is fixed,
regardless of the
actual material level.
Transducer acoustic beam
obstructed, standpipe too
narrow, or transducer ringing
(reads over 100%).
Relocate and / or re-aim trans-
ducer at material level or object.
Proceed to Measurement Diffi-
culties below.
See also: Transducer Ringing.
Material level
reported is always
incorrect by the same
amount.
Incorrect Empty (zero) refer-
ence for level operation
(P001 = 1).
See Empty (P006), Reading Off-
set (P063), Offset Calibration
(P650), & Offset Correction
(P652).
Measurement accu-
racy improves as
level nears trans-
ducer.
Incorrect Sound Velocity used
for distance calculation.
Use a transducer with a built-in
temperature sensor or a TS-3
temperature sensor.
See Sound Velocity.
Reading is erratic,
with little or no rela-
tion to material level.
True echo too weak or wrong
echo being processed.
Relocate and / or re-aim trans-
ducer at material.
Check noise parameters. See
Noise Problems.