Siemens HydroRanger 200 Manual de instrucciones Pagina 185

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 256
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 184
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 171
mmmmm
Parameters
P652 Offset Correction
The value altered when an Offset Calibration is performed.
Alternatively, if the amount of Offset Correction required is known, enter the amount to be
added to the Reading before display.
P653 Velocity
The value adjusted based on the Sound Velocity at 20 °C (P654) vs. Temperature (P664)
characteristics of air.
Alternatively, enter the current sound velocity (if known), or perform a Sound Velocity
Calibration (P651). The units used are m/s if P005 = 1, 2, or 3 (ft/s if P005 = 4 or 5).
P654 Velocity at 20°C
This value is used to automatically calculate Sound Velocity (P653).
After performing a Sound Velocity Calibration, check this value to verify the acoustic
beam atmosphere is air (344.1 m/s or 1129 ft/s).
Alternatively, if the acoustic beam atmosphere sound velocity at 20°C (68 °F) is known,
and the sound velocity vs. temperature characteristics are similar to that of air, enter the
sound velocity.
The units used are m/s if P005 = 1, 2, or 3 (or ft/s if P005 = 4 or 5).
Primary Index Global
Values Range: -999 to 999.0
Related P650 Offset Calibration
Primary Index Global
Values Range: 50.01 to 2001 m/s (164.1 to 6563 ft/s)
Related
P651 Sound Velocity Calibration
P654 Velocity at 20°C
Primary Index Global
Values Range: 50.01 to 2001 m/s (164.1 to 6563 ft/s)
Related
P005 Units
P651 Sound Velocity Calibration
P653 Velocity
Vista de pagina 184