Page 160 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01
mmmmm
Parameters
P603 Maximum Head
The level value associated with Maximum Flow, in Units (P005).
Represents the highest head level supported by the PMD and works in conjunction with
Maximum Flow (P604) to define the highest point in the exponential curve. Use it when
the PMD requires a maximum head and flow reference point. Includes Exponential,
Palmer Bowlus Flume, H-Flume, and Universal Breakpoints.
P604 Maximum Flow
The maximum flowrate associated with Maximum Head (P603).
Represents the flow at the highest head level supported by the PMD and works in
conjunction with Maximum Head (P603) to define the highest point in the exponential
curve. Use when the PMD requires a maximum head and flow reference point. Includes
Exponential, Palmer Bowlus Flume, H-Flume, and Universal breakpoints.
Also use this parameter with Time Units (P606) to define the flowrate units. The limitation
of four digits is for the LCD only, and the flowrate value is available with greater precision
through communications.
Example:
Primary Index Global
Values
Range: -999 to 9999
Preset: Span (P007) value
Altered By
• P005 Units
• P600 Primary Measuring Device (PMD)
Related
• P604 Maximum Flow
• P605 Zero Head
Primary Index Global
Values
Range: -999 to 9999
Preset: 1000
Altered By • P600 Primary Measuring Device(PMD)
Related
• P603 Maximum Head
• P606 Time Units
• P925 Flow Measurement
Conditions Enter
• Flowrate display: millions of gallons/day
• Max. flowrate is 376,500,000 gallons/day
• 376.5 for Maximum Flow (P604)
• 4 for Time Units (P606)