Page 176 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01
mmmmm
Parameters
Enter the time or distance interval over which the Rate Value is to be calculated before
the display updates.
This is automatically altered along with Maximum Process Speed (P003).
This value automatically alters the Rate Update Time (P705) and / or Rate Update
Distance (P706). Alternatively, these parameter values may be altered independently.
P705 Rate Update Time
The time period (in seconds) over which the material level rate of change is averaged
before Rate Value update.
P706 Rate Update Distance
The material level change (in metres) to initiate a Rate Value update.
P707 Rate Value
The rate of material level change (in Units (P005) or % of Span (P007) per minute).
A negative rate indicates the vessel is emptying.
This is the value displayed when is pressed in the RUN mode as described in the
Readings in RUN Mode
chart on page 18.
Alters • P707 Rate Value
Altered by • P003 Maximum Process Speed
Related • P705 Rate Update Time/ P706 Rate Update Distance
Primary Index Global
Values Range: 0.000 to 9999
Related • P707 Rate Value
Primary Index Global
Values Range: 0.000 to 9999
Related • P707 Rate Value
Primary Index Global
Values Range: -999 to 9999 (view only)
Altered By • P704 Rate Filter
Related • P005 Units • P007 Span