7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 163
mmmmm
Parameters
This parameter is enabled only if the primary measuring device (PMD) supports absolute
calculations (P600=6,7). For absolute PMDs (P600=6,7) volume units can be specified
using this parameter. If needed, absolute PMDs can still use ratiometric (P608=0) to
accommodate other units.
P610 Head Breakpoints
The head breakpoints for which flowrate is known.
The values in the Span for which flowrates are known. See
Universal Calculation Support
on page 76 for how to specify universal flows.
P611 Breakpoint Flowrates
The flowrate corresponding to each Head Breakpoint entered.
These are the flowrates for the related breakpoints. See
Universal Calculation Support
on
page 76 for how to specify universal flows.
Head vs. Flowrate (P610 and P611)
Primary Index Global
Secondary Index Breakpoint
Values Range: 0.000 to 9999
Related • P611 Breakpoint Flowrates
Primary Index Global
Secondary Index Breakpoint
Values Range: 0.000 to 9999
Related • P610 Head Breakpoints
Max. Flow
Zero Flow
Zero Head
Max. Head
Head Breakpoints
Flow Breakpoints