Site Considerations, Continued
When the Maxum II contains an air bath oven, it will vent in excess of 3 SCFM (85 liters per minute) of
heated air through the oven exhaust. See note below for the recommended approach.
WARNING All vent lines and oven exhaust should be vented outside the shelter. The vents
should be connected to a location where the pressure is atmospheric. Connection of oven or
analyzer vents to a variable pressure source such as a flare header should not be done, as
chromatographic performance will be severely affected. Backpressure on the oven vent could
cause the T-rating of the instrument to be exceeded, which might become an ignition source.
Refer to the comments on Hazardous gases on page 3.
Installation of the unit with a combustible gas coming into the EPC requires that a restriction be placed on
the flow into the Maxum II. This is required to limit the flow of the combustible gas into the unit. The
recommendation is to limit the flow prior to entry into the shelter.
WARNING The Maxum II will be tagged for flow restriction of combustible gases into the
Electronic Pressure Controller. The user/installer must ensure that this flow restriction is
implemented to permit safe operation of the unit.
Intended Use
The Maxum edition II gas chromatograph is used in all branches of the fine chemicals, refining and
hydrocarbon processing industries. It performs chemical composition analysis of gases and liquids that
are present in all phases of production. The Maxum II is built for installation in harsh environments either
directly or nearby in at-line process measurement laboratories. Its application flexibility allows it to
analyze samples of feedstock, partially processed streams, final products and process byproducts
including wastes and environmental hazards.
This product is intended to be used only in conjunction with other devices and components which have
been recommended and approved by Siemens. Appropriate safety standards were used in the
development, manufacture, testing, and documentation of the Maxum II. Under normal operation, this
product is safe for use providing that all safety and handling guidelines are observed with respect to
configuration, assembly, approved use, and maintenance. This device has been designed such that safe
isolation is guaranteed between high and low voltage circuits. Low voltages which are connected must
also be generated using safe isolation.
If any part of the Maxum II is opened, certain parts of the device are accessible which may carry
dangerous voltages. Therefore, only suitably qualified personnel may work on this device as indicated in
the next section which is titled “Qualified Personnel”.