Sampling System Installation, Continued
Figure 2-4. Typical Installed Sample Probe
Returning Samples to the Process Stream
Refer to the Application Drawing Package, Sampling System sheet and Block Diagram Utility
Connections sheet.
If the sample stream is a gas and is to be released to the atmosphere, install a vent line outside the
analyzer house to at least 8 ft. (2.4 m) above the ground. The exhaust end should be curved down, or a
vent cap added for weather protection. If the sample gas is to be returned to a pressurized line, a
minimum pollution valve is required in the sampling system. The return line should have only a very low
pressure (0 to 20 psig [0 to 138 kPa].) Install a check valve to prevent backflow.
When returning a gas sample that can condense, slope it downward to a liquid drain or return it to the
process stream using heat tracing to prevent any condensation. For details on installing heat tracing,
refer to the "Sampling System" description in this section.