Maxum II Specifications, Continued
I/O Options
Standard I/O 2 analog outputs; 4 digital outputs (1 indicates system error, 3 are
user configurable); 4 digital inputs; 1 serial output
Board Slots for Optional I/O 2
I/O Boards AO 8: 8 electrically isolated analog output channels
D I/O: 4 digital inputs and 4 digital outputs
A I/O: 2 digital inputs and 2 digital outputs,
2 analog inputs and 2 analog outputs
Digital Inputs Optocoupler with internal 12-24 Vdc power supply, switchable
with floating contacts. Alternative: switchable with external 12-24
Vdc supply (floating relay contacts only), external power supply
negative terminal connected to common for specific DI being used
Digital Outputs Floating double-throw contacts, maximum contact load rating
1 A at 30 Vdc. Diode shunt suppression should be used for
inductive loads. A DO should not drive more than a 0.5A
inductive load. Dry contact DOs (I/O boards and Syscon) can not
be used with line voltages. The PECM has SSRs which may be
controlling line voltage for the particular application
Analog Inputs -20 to +20 mA into 50 ohms or -10 to +10 V R
in
= 1 M ohm,
mutually isolated to 10 V
Analog Outputs 0/4 to 20 mA into 750 ohms max., common negative pole,
galvanically separated from ground, freely connectable to ground
Termination Screw compression terminal strip for braided or solid cable with
maximum section of 16 AWG or 1.5 mm
2
Gas Sample Requirements (application dependent)
Sample Flow 50-200 cc/min (application dependent)
Sample Filtration 0.1 μm
Minimum Sample Pressure 5 psig (35 kPa), lower pressure optional
Maximum Sample Pressure 75 psig (515 kPa), standard; higher pressure optional
Maximum Sample Temperature 250°F (121°C) standard; higher temperature optional
Material in Contact with Sample Stainless steel, Teflon, and polyimide; other material optional
Liquid Sample Requirements (application dependent)
Sample Flow 5-20 cc/min (application dependent)
Sample Filtration 5 μm
0.3 μm if SLIV used in application
Minimum Sample Pressure 5 psig (35 kPa), standard
Maximum Sample Pressure 300 psig (2070 kPa) standard; higher pressure optional
Maximum Sample Temperature 250°F (121°C) standard; higher temperature optional
Material in Contact with Sample Stainless steel and Teflon; other material optional