Receiving a Crated Analyzer
Instructions
When the Maxum II is shipped to the installation site in a reinforced crate with vibration protection, it will
be clearly labeled for horizontal placement and handling.
On receipt of goods, look for any visible damage and verify that all items noted to be shipped
were received. Record on the goods receipt any damage or missing items noting the item(s) and
quantity missing on all copies. If anything is damaged or missing, then also notify your Siemens
representative or authorized agent of the damage to the equipment.
CAUTION If the unit is to be stored, it should be protected from the weather (the crate does not
provide weather protection.) If left outside in the weather, severe damage to the unit may occur. If
the unit is to be shipped to another location, preserve the vibration isolation provided in the
original shipping crate, or address the vibration isolation in the shipment method.
WARNING A crated Maxum II will weigh a minimum of 200 pounds (90 kg), and can weigh
substantially more depending upon the configuration. It should be moved with a forklift until
uncrated.
Place the crate on the floor horizontally with the orientation
label THIS SIDE UP pointed up.
Remove the metal bands from around the crate.
Unscrew the cover and remove the crate cover. Remove
the crate sides to facilitate removal of the unit. Remove the
2x4 supports and foam blocks from the top of the analyzer.
The analyzer will be covered in a plastic sheet.
The Maxum II will be bolted to one wooden support, which
should remain on the analyzer until it reaches its final
mounting location. Do not remove the bolts holding the
Maxum II to the wooden supports. Remove the bolts holding the mounting brackets to the base of
the unit. Leave them on the analyzer for use as handholds, skids, or lifting bars.
CAUTION The Maxum II must be lifted only by the mounting brackets on the back. Never allow
the analyzer to be supported on its side or front. Lifting with of any of the equipment protruding
from the analyzer will damage the equipment.
Remove the Maxum II from the shipping crate.
Remove the plastic bag and drying agent from around the analyzer.
Examine the Maxum II for visible damage. Notify the carrier of the damage and file the necessary
claims. Also notify your Siemens representative or authorized agent of the damage to the
equipment.