Analyzer Connections, Continued
For an analyzer with an FID, connect a length of new, clean, ¼” (6.35 mm) stainless steel tubing
to the analyzer port labeled "FLAME AIR OUT," and run it to the inlet port on a separately
mounted air treater.
For an analyzer with an FID, connect a length of new, clean, ¼” (6.35 mm) stainless steel tubing
to the instrument air supply, and run it to the outlet port on the separately mounted air treater.
Mount the air treater in a convenient location near the analyzer. The mounting bracket requires
four ¼” (6.35 mm) diameter bolts.
The air treater can be mounted on a wall in whichever direction is best for connecting the ¼”
(6.35 mm) tubing to and from the analyzer.
The analyzer oven exhaust vent(s) should be routed outside the building or shelter. See warning
concerning vent lines on page
30.
Remove the plugs from the sample supply lines. If Siemens is supplying the sampling system, the
sample conditioning portion will probably be mounted in a cabinet or on a panel beneath the
analyzer oven. Connect the sample supply lines to the port labeled by their stream numbers on
the side of the cabinet or on the panel; refer to the Application Drawing Package, Sampling
System Connections sheet.
Remove the plug from the calibration standard supply line. Connect the line to the port labeled
"STANDARD" on the sampling system cabinet or panel; refer to Application Drawing Package,
Sampling System Connections sheet.
Connect drain, or return lines, as required to the appropriate sampling system vent refer to
Application Drawing Package, Sampling System Connections sheet.