Maxum II Specifications, Continued
AC Power
100-130 VAC or 187-264 VAC (configuration dependent); 47-63 Hz.,
single phase
Transient Overvoltage – Category II (IEC 60364-4-443)
Single oven: 14 amperes maximum.
Dual oven: 2 circuits, 14 amperes maximum per oven
To meet European Standards, customer provided power cables
should be double insulated (unless both power and signal wiring are
rated for 300 volts, in which case basic insulation is sufficient)
Instrument Air 50 psig (350 kPa) minimum for units using Model 11 or Valco valves
120 psig (825 kPa) minimum for units using Model 50 valves
Air bath: 25 psig (175 kPa) minimum; 3 scfm (85 liters per
minute)/oven
Airless: None for heating (may be used for electronics purge)
Carrier Gas Cylinder nitrogen or helium at 99.998% purity, or hydrogen at
99.999% purity depending on application
Typical consumption – 180 scf/month/detector module
(5100 liters/month/detector module)
Flame Fuel Hydrogen at 99.999% purity with no more than 0.5 ppm total
hydrocarbons
Typical consumption – 70 scf/month/detector module
(2000 liters/month/detector module)
Flame Air Zero grade air (< 1ppm THC, O
2
content 20-21%). Supplied from
instrument air with catalytic purifier (optional). Typical consumption –
900 scf/month (26,000 L/month)
Corrosion Protection Stainless steel oven liner
Painted steel exterior (epoxy powder coat)
Dry air purge to protect electronics. To maintain optimum
performance, the analyzer electronics must be kept clean and dry. If
the atmosphere in the installation environment contains condensing
liquids or corrosive gases, the electronics compartment should be
purged with clean, dry air
Calibration
Type Manual or automatic
Zero Automatic baseline correction
Span Standard sample cylinder