UM344-2 INITIAL TRANSMITTER SET-UP
September 1995
3
-
7
TEST SCRN - Press to test all screen segments.
When this key is pressed, the MXC displays the
"Begin" message to the left, then lights all the
segments, turns them all off, displays the
"Completed" message, and then returns to the
original test screen.
END - Press to exit the test mode and return to the
Main Menu screen.
3.3.2 Transmitter Testing
Although the Transmitter continuously performs an on-
line self test, a more extensive self test can be performed
after communication with the MXC has been established.
1. At the On-Line Menu, press CAL/TEST (F2) to
display the Calibrate and Test selections.
2. At the Calibrate/Test Menu, press TEST. A Test
caution screen will appear. Press one of the
following:
ABORT - Press to return to the Calibrate/Test
Menu.
PROCEED - Press to continue the self test. A
warning screen will be displayed then a message
states that the test is occurring. After the test, the
MXC will show whether or not the transmitter has
passed or failed the self test.
If the MXC indicates that the Transmitter has:
Passed - Press CONT then END.
Failed - Check installation wiring thoroughly. Go to
Section 6 for troubleshooting suggestions.
3. When the MXC displays the Transmitter’s ID (On-
Line Menu), press the STATUS Key to check for
Transmitter errors. If an error is present, the
ERROR Key (F2) will be displayed. Press
ERRORS (F2) to check for ROM, RAM, EEPROM,
TIMER and SENSOR errors. Refer to Section 6.3
Troubleshooting to confirm and resolve the error.
4. If the alarm function is enabled, press ALARM (F1)
to check the ALARM OUT OF SERVICE status.
MXC SCREEN TEST
ABOUT TO BEGIN
MXC SCREEN TEST
COMPLETED
TEST - PERFORMS
DIAGNOSTIC TEST ON
TRANS- A- PRO-
MITTER BORT CEED
CALIBRATE / TEST
CAL TEST END
WARNING! SELF TEST
MAY BUMP TRANSMITTER
OUTPUT A-
BORT CONT
TRANSMITTER PASSED
TRANSMITTER SELFTEST
CONT