4.2 Disassembly
WARNING
Incorrect disassembly
The following dangers may result through incorrect disassembly:
- Injury through electric shock
- Danger through emerging media when connected to the process
- Danger of explosion in hazardous area
In order to disassemble correctly, observe the following:
Before
starting work, make sure that you have switched off all physical variables such as
pressure, temperature, electricity etc. or that they have a harmless value.
If the device contains dangerous media, it must be emptied prior to disassembly. Make
sure that no environmentally hazardous media are released.
Secure the remaining connections so that no damage can result if the process is started
unintentionally.
4.3 Installation (except level)
4.3.1
Instructions for installation (except level)
Conditions
Note
Compare the desired operating data with the data on the nameplate.
Please also refer to the information on the remote seal if this is fitted.
Note
Protect the transmitter against:
Direct heat radiation
Rapid temperature fluctuations
Heavy contamination
Mechanical damage
Direct sunlight
The installation location is to be as follows:
Installing/mounting
4.3 Installation (except level)
SITRANS P, DS III series with HART
Operating Instructions, 09/2012, A5E00047092-08 39