Flange connection in accordance with EN 61518
Flush-mounted process connections
3.6.3.1 Measuring cell for gauge pressure
S
H

Reference pressure opening Filling liquid
Measuring cell Gauge pressure sensor
Process connection p
e
Inlet pressure
Seal diaphragm
Figure 3-7 Function chart of measuring cell for gauge pressure
The
inlet pressure (p
e
) is transferred to the gauge pressure sensor via the seal diaphragm
and the fill fluid , displacing its measuring diaphragm. The displacement changes the
resistance of the four piezoresistors (bridge circuit) of the gauge pressure sensor. The change
in the resistance causes a bridge output voltage proportional to the inlet pressure.
Transmitters with measuring span ≤ 63 bar measure the inlet pressure against atmosphere,
those with measuring spans ≥ 160 bar against vacuum.
Description
3.6 Functional principle
SITRANS P, DS III series with HART
Operating Instructions, 09/2012, A5E00047092-08 25