Example of an application
The
current is to be measured as a voltage drop from 1 to 5 V at a resistance of 250 Ohm +/-
5%. To trim the tolerance of the resistance, set the current transmitter so that the voltage drop
at 4 mA is exactly 1 V and at 20 mA is exactly 5 V.
● Trim at 4 mA:
Use the current transmitter trim menu option to instruct the transmitter to output 4 mA. Read
the measured value at an ammeter and input it. The transmitter uses this value for offset
correction of the current.
● Trim at 20 mA:
Use the current transmitter trim menu option to instruct the transmitter to output 20 mA.
Read the measured value at an ammeter and input it. The transmitter uses this value for
gradient correction of the current. The value for 4 mA is not affected by this.
Note
If a multimeter is used, it must always be sufficiently accurate.
$%
,
,
P$
P$
%
&
%&
$
0HDVXUHG
FXUUHQW,
5DWHGFXUUHQW,
9DULDWLRQLQULVH
3DUDOOHOVKLIW
A Original characteristic
B Characteristic curve after current transmitter trim 4 mA
C Characteristic curve after current transmitter trim 20 mA
Figure 7-15 Current transmitter trim
7.18
Factory calibration
Introduction
You can use factory calibration to reset the transmitter to the factory state.
Operating functions through HART
7.18 Factory calibration
SITRANS P, DS III series with HART