Function
The input pressure is converted into an electrical signal by the sensor .
This signal is amplified by the measuring amplifier and digitized in an analog-to-digital
converter .
The digital signal is analyzed in a microcontroller and corrected with regard to linearity
and thermal characteristics.
The digital signal is then converted in a digital-to-analog converter into the output current
of 4 to 20 mA. A diode circuit provides reverse voltage protection.
You can make an uninterrupted current measurement with a low resistance ammeter at
the connection .
The measuring cell-specific data, electronics data and parameter assignment data are
saved in two EEPROMs . The first memory is linked to the measuring cell, the second
to the electronics.
Operation
The buttons can be used to call up individual functions, so-called modes.
If you have a device with a display , you can use it to track the mode settings and other
device messages.
The basic mode settings can be changed with a computer and HART modem via PDM.
3.6.3 Measuring cell operation
CAUTION
If the measurement signal fails because of sensor breakage, the seal diaphragm may also
be destroyed. In the worst case scenario, the process medium leaks from the reference
pressure
opening in the devices used for gauge pressure with a measuring span of ≤ 63 bar.
In the following sections, the process variable to be measured is called general inlet pressure.
Overview
The following modes of operation are described:
Gauge pressure
Absolute pressure
Differential pressure and flow rate
Level
The following process connections are available, for example:
G1/2 B, 1/2-14 NPT
Male thread: M20
Description
3.6 Functional principle
SITRANS P, DS III series with HART
24 Operating Instructions, 09/2012, A5E00047092-08