NOTICE
Electrostatic-sensitive devices
The
device contains electrostatic-sensitive devices (ESD). ESD can be destroyed by voltages
far too low to be detected by humans. These voltages can occur if you simply touch a
component part or the electrical connections of a module without being electrostatically
discharged. The damage to a module caused by overvoltage cannot normally be detected
immediately; it only becomes apparent after a longer period of operating time has elapsed.
Protective measures against the discharge of static electricity:
● Make sure that no power is applied.
● Before working with modules, make sure that you discharge static from your body, for
example by touching a grounded object.
● Devices and tools used must be free of static charge.
● Hold modules only by their edges.
● Do not touch connector pins or conductor tracks on a module with the ESD notice.
10.2 Maintenance and repair work
10.2.1
Defining the maintenance interval
WARNING
No maintenance interval has been defined
Device failure, device damage, and risk of injury.
● Define
a maintenance interval for regular tests in line with device use and empirical values.
● The maintenance interval will vary from site to site depending on corrosion resistance.
Service and maintenance
10.2 Maintenance and repair work
SITRANS P, DS III series with HART
162 Operating Instructions, 09/2012, A5E00047092-08