Additional Information CLV61x Dual Port (PROFINET IO)
8017978 / 2014-11-20 © SICK AG · Germany · All rights reserved · Subject to change without notice 45
7.8 Firmware update of the CLV61x Dual Port via SOPAS ET
The CLV61x Dual Port be updated with SOPAS ET via USB or via TCP/IP (Port 2111 and 2112). If the
update is carried out via Ethernet, the IP address must be accessible in a stable condition, i.e., the CLV61x
Dual Port must retain this IP address even after a restart, or retrieve it from the attached PLC.
It may be necessary to test beforehand whether the CLV61x Dual Port is reachable after rebooting.
1. Use the SOPAS ET Network Scan Assistant to scan the network
The CLV61x Dual Port is found via USB
2. Add attached ID sensor to the project:
3. Log into the device using the password for user-level service: "servicelevel":
5. Use the command "CLV ... > Firmware Download" to open the download window:
6. Select the current firmware (.fwp file) in the download window and the matching *.key file:
7. Start download. Password for user-level service: "servicelevel":
8. Depending on the ID sensor, the download can take up to approx. 40 min. and the message "Download
succeeded" will appear once it is complete:
Vista de pagina 44