SITRANS P measuring instruments for pressure
Transmitters for gage, absolute and differential pressure
Z series for gage pressure
2/5
Siemens FI 01 · 2009
2
■
Technical specifications
■
Dimensional drawings
SITRANS P pressure transmitters, Z series (7MF1562-...), dimensions in
mm (inch)
■
Schematics
SITRANS P pressure transmitters, Z series (7MF1562-...), connection dia-
gram
D) Subject to export regulations AL: N, ECCN: EAR99H.
SITRANS P pressure transmitter, Z series for gage pressure
Mode of operation
Measuring principle Thin-film strain gage
Input
Measured variable Realtive pressure
Measured range 0 to 16 bar g (0 to 232 psi g) or
0 to 25 bar g (0 to 363 psi g)
Output
Current output signal 4 ... 20 mA
Measuring accuracy To EN 60770-1
Error in measurement (at 25 °C
(77 °F), including conformity error,
hysteresis and repeatability)
0.5% of full-scale value-typical
Response time T
99
< 0.1 s
Long-term drift
• Start of scale 0.3% of full-scale value/year - typ-
ical
• Measured span 0.3% of full-scale value/year - typ-
ical
Influence of ambient temperature
• Start of scale 0.3%/10 K (0.3%/10 K) of full-
scale value - typical
• Measured span 0.3%/10 K (0.3%/10 K) of full-
scale value - typical
Rated conditions
Medium conditions
• Process temperature -30 ... +120 °C (-22 ... +248 °F)
Degree of protection to EN 60529 IP65
Ambient conditions
• Ambient temperature -25 ... 85 °C (-13 ... +185 °F)
• Storage temperature -50 ... 100 °C (-58 ... +212 °F)
Design
Weight ≈ 0.2 kg (≈ 0.44 lb)
Wetted parts materials
• Measuring cell Al
2
O
3
- 96%
• Process connection Brass, mat. No. 2.0402
• Gasket Viton
Process connection Male thread G½B
female thread G
1
/
8
B
Power supply
Terminal voltage on pressure trans-
mitter
• For current output 10 to 36 V DC
Certificate and approvals
Classification according to pressure
equipment directive
(DRGL 97/23/EC)
For gases of fluid group 1 and
liquids of fluid 1; complies with
requirements of article 3, para-
graph 3 (sound engineering prac-
tice)
6:
G
/
8
B
*%
0[
ෘ
GLDP
1+
2-
+
Signal
R
L
U
B
I
0
I
0
Output current
U
B
Power supply
R
L
Load
Connections:
1 (+U
B
)
2 (-U
B
)
■
Selection and Ordering data Order No. Order code
SITRANS P pressure transmitters, Z series for pressure
2-wire system, characteristic rising
D)
7MF1562 -
777
0 0
7 7 7
Measured range Max. working pressure
0 ... 16 bar g (0 ... 232 psi g) 32 bar g (464 psi g) 3CB
0 ... 25 bar g (0 ... 363 psi g) 64 bar g (928 psi g) 3CD
Other version for measuring range ≥ 1 bar g (≥ 14.5 psi g), add Order code and plain text:
Measuring range: ... to ... bar g (psi g)
9AA H 1 Y