SITRANS P measuring instruments for pressure
Remote seals
Measuring setups
2/182
Siemens FI 01 · 2009
2
Overview
This section shows examples of typical measuring setups for us-
ing SITRANS P pressure transmitters with and without remote
seals.
Equations for calculating start of scale and full scale are pro-
vided for each example.
Questionnaires are included to help you select the right combi-
nation of remote seal and pressure transmitter.
Installation
Remote seals of sandwich design are fitted between the con-
nection flange of the measuring point and a dummy flange. Re-
mote seals of flange design are fitted directly on the connection
flange of the measuring point. The r
espective pressure rating of
the dummy flange or the flanged remote seal must be observed.
The pressure transmitter should be installed below the connec-
tion flange (and below the lower connection flange in the case of
differential pressure transmitters). This arrangement must
be
used in the low-pressure range.
When measuring at pressures above atmospheric, the pressure
transmitter can also be installed above the connection flange.
The capillaries between the remote seal and the pressure trans-
mitter should be as short as possible to obtain a good transmis-
sion response.
Offset of measuring range
If there is a difference in height between the two connection
flanges when measuring with two remote seals, an additional dif-
ferential pr
essure will result from the oil filling of the remote seal
capillaries. This results in a measuring range offset which has to
be taken into account when you set the pressure transmitter.
An offset in the measuring range also occurs when combining a
remote seal with a transmitter if the remote seal is not installed at
the same height as the transmitter.
Pressure transmitter output
If the level, separation layer or density increase in closed ves-
sels, the dif
ferential pr
essure and hence the output signal of the
pressure transmitter also increase.
For an inverted relationship between the differential pressure
and the output signal, the start-of-scale and full-scale values of
the SITRANS P must be interchanged.
With open vessels, a rising pressure is usually assigned to an in-
creasing level, separation layer or density.
Influence of ambient temperature
Temperature differences between the individual capillaries and
between the individual r
emote seals should be avoided.
Temperatur
e variations in the area of the measuring setup cause
a change in volume of the filling liquid and hence measuring er-
rors.
Notes
For the separation layer measurement, the separation layer
has to be positioned between the two spigots. Also you must
make sur
e that the level in the container is always above the
top spigot.
W
hen measuring density, make sure that the level of the me-
dium remains constant. The level should be above the top
spigot.
Possible combinations of pressure transmitters and remote
seals
Type of installation Pressure transmit-
ters
Remote seals
A/B 7MF4033
7MF4034
7MF4035
7MF8023
7MF8024
7MF8025
7MF4900
7MF4910
7MF4920
C
1
and C
2
7MF4233
7MF4234
7MF4235
7MF4900
7MF4910
7MF4920
(vacuum-proof design
in each case)
7MF4333
7MF4334
7MF4335
7MF4901
7MF4921
D 7MF4433
7MF4434
7MF4435
7MF4903
7MF4923
E 7MF4433
7MF4434
7MF4435
7MF4913
G, H and J 7MF4433
7MF4434
7MF4435
7MF4903
7MF4923
© Siemens AG 2008