IND sensors
  
Siemens PA 20 · 1999
1/34
Technical data
IND sensor 7MA2ā200 - . . -8BA -8BF -8CB -8DA -8DD -8BD -8EA -8EB
Cell constant cm
Ć1
3.82 3.16 3.00 5 3.82 3.82
Min. measuring range/
max. measuring range
0 to 0.1 mS/cm/
0 to 2500 mS/cm
1
)
BuiltĆin thermometer No Yes No Yes
Resistance thermometer 90% time Ć 100 s 50 s 100 s Ć 100 s
Wetted parts material FEP
2
) PEEK
3
) DURAN
4
) FEP
2
)
Type of connection
Conical flange
DN 50
VARIVENT
Con. flange
DN 50
VARIVENT
Flange
ISO 3587
Conical flange
DN 50
Permissible oper. pressure P
B
at T
B
10 bar 4 bar 10 bar
Permissible oper. temperature T
B
for immersion part
130 ºC
Permissible storage temperature
Ć25 to +85 _C
Fixed cable 5m
Max. recommended cable length
(with max. measuring range)
< 50 m
Type of cable 3x2x0.25 LIYCYĆCY
Ex protection to DIN 50014/EN 50020 _
II 2G EEx ib IIC T4
5
)
zone 1
Max. cable length with Ex protection _
Siemens cable
C79451ĆA3300ĆN100,
max. 5 m long
Degree of protection to DIN 40050 IP 65 IP 67
Weight Approx. 1.2 kg
Dimensions See Fig. 1/25
1
) When using the panel housing, smallest measuring range 0 to 2 mS/cm
2
) FEP: perfluoroethylenepropylene
3
) PEEK: polyetheretherketone
4
) DURAN: special glass
5
) When using SIPAN 32 Ex analyzer
% w/w
40
60
100
20
80
110
10 20 30 40 80 50 1004030
40
60
100
20
80
90 100 20
110
T/_C
T/_C
% w/w
HCl
NaOH H
2
SO
4
H
2
SO
4
HNO
3
KOH
HBr
HNO
3
Fig. 1/28 SIPAN 32, SIPAN 32X and SIPAN 34 analyzers, ranges in which conversion into % w/w is not physically possible