2 EL sensors, 4 EL sensors
Siemens PA 20 · 1999
1/33
Technical data
2EL sensor 7MA2000 - . . -8AA -8AB -8BA -8BB -8CA -8CB -8DD -8DS
Cell constant cm
Ć1
0.00580 0.0580 1.160 0.10 0.50
Min. measuring range/
max. measuring range S/cm
0 to 0.1 /
0to25
0 to 125
3
)
0to1/
0 to 250
0 to ā1250
3
)
0to20/
0 to 5ā000
0 to 25ā000
3
)
0to2/
0to500
0 to 2500
3
)
0to 10 /
0 to 2ā500
0 to 125ā00
3
)
Resistance thermometer 90% time 60 s 3s 20 s
Wetted parts material Stainless steel (1.4571) + glass
Electr. carbon + glass
Stainl. steel (1.4571)
PES
1
)
+
stainl. steel
(1.4571)
PPO
2
)
+
stainl. steel
(1.4571)
Mounting
6Ćhole
flange
Union
nut
6Ćhole
flange
Union
nut
6Ćhole
flange
Union
nut
Union nut Pg 13.5
Type of connection Conical flange DN 50 Pg 13.5
Permissible oper. pressure P
B
at T
B
16 bar 6 bar 2 bar
Permissible oper. temperature T
B
130 ºC 60 ºC 80 ºC
Permissible storage temperature
Ć25 to +85 _C
Cable Ć 5 m fixed
5 m plugged
LIYCY 4 x 0.14
Type of cable 3x2x0.2LIYCY Ć
Max. recommended cable length
(with max. measuring range)
< 50 m
Ex protection DIN 50014/EN 50020 In conjunction with SIPAN 32X, all sensors are approved for use in Ex zone 1
Max. cable length with Ex protection Siemens cable C79451ĆA3298ĆN100, max. 5 m long
Degree of protection to DIN 40050 IP 54 IP 65
Weight Approx. 1.5 kg Appr. 0.5 kg Appr. 0.2 kg
Dimensions See Fig. 1/21
4EL sensor 7MA2100 - . . -8BC -8CA
Cell constant cm
Ć1
0.0471 0.0828
Min. measuring range/
max. measuring range in mS/cm
0 to 0.1 /
0to500
Resistance thermometer 90% time 100 s 40 s
Wetted parts material Epoxy resin with graphite
Conical flange DN 50 Yes No
Mounting Union nut Pg 13.5
Permissible oper. pressure P
B
at T
B
6 bar
Permissible oper. temperature T
B
Continuous
ShortĆterm
100 _C
110
_C
70 _C
110
_C
Permissible storage temperature
Ć25 to +85 _C
Fixed cable Ć 5m
PlugĆtype cable 5m Ć
Max. recommended cable length
(with max. measuring range)
< 50 m < 50 m
Type of cable 4x2x0.2LIYCY
Ex protection DIN 50014/EN 50020
In conjunction with SIPAN
32X, all sensors are
approved for use in
Ex zone 1
Max. cable length with Ex protection
Siemens cable
C79195ĆA3453ĆN100,
max. 5 m long
Degree of protection to DIN 40050 IP 54 IP 65
Weight Appr. 1 kg Appr. 0.5 kg
Dimensions See Fig. 1/23
1
) PES: polyethersulfone;
2
) PPO: polyphenylene oxide
3
) Reduced measuring accuracy