Siemens ET 78X Series Manual de usuario Pagina 18

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 38
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 17
Characteristics
  
Siemens PA 20 · 1999
1/18
SIPAN 34 analyzer
SIPAN 34 is an analyzer of the fourĆwire generation with
stateĆofĆtheĆart technology with microprocessor control and
illuminated graphic display.
The SIPAN 34 analyzer is optionally available with special
features for process use.
The SIPAN 34 analyzer is available in two designs:
- With a field housing
- With a panel housing
It contains the analog and digital data processing functions for
the signal delivered by the sensor.
A SIPAN 34 analyzer can be used for all measuring ranges.
Special characteristics of SIPAN 34
FourĆwire analyzer with extremely easy operation
Universal power supply (24 V AC/DC, 115 V AC, 230 V AC)
Complete basic configuration
SelfĆexplanatory menu operation in plain text in five
languages (German , English, French, Spanish, Italian), help
function
Operation according to NAMUR, i.e. complete local
operation with directly accessible keypad with 8 keys and
large, illuminated, fullĆgraphic display
Display of S/cm, mS/cm, mS/cm, mS/m, MWcm, kWcm,
%w/w,H
2
SO
4
, oleum, HNO
3
, HCl, HBr, NaOH, NaCl, KOH
Direct output of concentration values instead of conductivity
(19 stored material tables)
Additional permanent bargraph of measuring range
Graphic trend display of measured value
Additional permanent temperature display in °C
Output signal 0/4 to 20 mA, floating
FreelyĆprogrammable, permanent measuringĆpoint
designation (saves tag labels)
Logbook with entry of faults or calibration procedures with
date and time
Fault and limit contacts
NonĆlinear ultraĆpure water temperature compensation for
conductivity
All conductivity measuring procedures (2EL/4EL/IND) can be
selected, i.e. only one analyzer is required for the complete
conductivity range
Program for automatic recording of userĆspecific temperature
compensation
Maintenance switch with automatic HOLD function
Comprehensive fault diagnosis and preventive maintenance
system in plain text
3 operating levels with coded protection for monitoring,
routine and specialists
Selectable tests for: keys, RAM, EPROM, EEPROM, display
Output of freelyĆdefined current values for test purposes
Maximum electromagnetic compatibility according to CE and
NAMUR, sensitive lightning protection
Panel housing made completely of metal, CE safety for every
control cabinet installation engineer
Robust field housing (IP 65) with seven Pg screwed glands
for easy connection
No special or expensive mounting set required for wall or
panel mounting
Options for SIPAN 34
Second current output for measured value or temperature
with additional limit
Four parameter sets with remote selection for complete
methods, not only for measuring ranges, e.g. also limits,
physical dimensions, temperature compensation with
complete characteristic (not only TC value), hysteresis
Individual calibration of each parameter set possible
Automatic cleaning function (3 relays) for cleaning, flushing,
fitting control with cyclic time input, waiting and holding
functions
TwoĆpoint controller for pulse length (dosing valves) or pulse
frequency (diaphragm pumps)
Additional switching contact for maintenance (function
check) and preĆalarm (warning)
Functions
Basic analyzer Options
Inputs
Conductivity
Temperature
Remote switching of range for 4
parameter sets, thus access to 4
complete parameter sets for complete
methods including ranges, limits,
physical dimensions, temperature
compensation, hysteresis
Outputs
Analog output 2nd analog output for temperature
Contacts
1x failure
1x limit and
2x NAMUR contacts
2 limits
with
controlĆ
ler
function
Second
limit
3 x cleaning or
3 x range
signalling
contacts
or
Vista de pagina 17