7ML19985BE02 SITRANS LU01 – INSTRUCTION MANUAL Page 47
mmmmm
Parameters
mA Output (P200 to P219)
When an mA Output Parameter is accessed, the mA symbol is displayed in the Index Type
field and the mA output number (corresponding to the SITRANS LU01 terminals) is
displayed in the Index field.
P200 mA Output Range
Determines the mA output range.
If either 1 or 2 is selected, the mA output is directly proportional to the mA Function. If
either 3 or 4 is selected, then the output is inversely proportional.
P201 mA Output Function
Alters the mA output/measurement relationship.
P203 mA Output Value / Transducer
Displays current mA output value.
This displays as an Auxiliary Reading when key is pressed in the RUN mode and
does not include adjustments made using Trim features (P214 / P215).
Primary Index mA output
Values
0off
1 0 to 20 mA
20 mA = high level (unless P001 = 2
or 3, then 20 mA = low)
2 * 4 to 20 mA
3 20 to 0 mA
20 mA = low level (unless P001 = 2
or 3, then 20 mA = high)
4 20 to 4 mA
Related • P911 mA Output Value
Primary Index mA output
Values
value mA function Operation (P001)
0OFF
1level level
2space space
3 distance distance
4 volume
Related
• P202 mA Output Allocation
•P911 mA Output Value
Altered By •P001 Operation
Values Range: 0.000 to 22.00 (view only)
Note: This parameter is applicable only if any mA output has the transducer
Index Number as its input source (see P201).
Comentarios a estos manuales