7ML19985BE02 SITRANS LU01 – INSTRUCTION MANUAL Page 35
mmmmm
Parameters
Entering a Level Breakpoint
1. Go to Parameter P054.
2. Press to display the index symbol.
3. For each index enter a volume.
4. Ensure that each volume corresponds to the same index for P055.
5. Press .
For more on Volume Characterization, go to page 87.
P055 Volume Breakpoints and Characterization
(Universal Volume Calculation)
Each segment defined by the level breakpoints (P055) requires a volume so that the
SITRANS LU01 can make the level-to-volume calculations.
Typical volume calculations
Entering a Volume Breakpoint
1. Go to Parameter P055.
2. Press to display the index symbol.
3. For each index enter a volume.
4. Ensure that each volume corresponds to the same index for P054.
5. Press .
For more on Volume Characterization, go to page 87.
Primary Index Breakpoint
Values Range: 0.0 to 9999
Related • P054 Breakpoints Levels (Universal Volume Calculation)
Cone Cylinder
V = (1/3)πr
2
hV = πr
2
h