Siemens SITRANS LR 460 Manual de usuario Pagina 12

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 123
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 11
Page 6 SITRANS LR 460 – INSTRUCTION MANUAL 7ML19985JM01
mmmmm
Specifications
Specifications
SITRANS LR 460
Power
Power Supply
100 to 230 V AC,
±15%, 50/60 Hz, 6 W
24 V DC, +25/-20%, 6 W
Fuse (AC) SI1 Fast acting ceramic, 4 x 20 mm, 1 A, 250 V AC
SI2 Slow-Blow, 4 x 20 mm, 0.63 A, 250 V AC
Fuse (DC) SI1 Fast acting ceramic, 4 x 20 mm, 2 A, 250 V AC
SI2 Slow-Blow, 4 x 20 mm, 0.63 A, 250 V AC
Performance
Reference operating conditions according to IEC 60770-1
ambient temperature +15 to +25
o
C
humidity 45 to 75% relative humidity
ambient pressure 860 to 1060 mbar (12.47 to 15.37 psi)
Measurement Accuracy (measured in accordance with IEC 60770-1)
non-linearity (accuracy) greater of 25 mm (1”) or 0.25% of span (including
hysteresis and non-repeatability)
non-repeatability 10 mm (0.4”) [included in non-linearity specification]
deadband (resolution) 10 mm (0.4”) [included in non-linearity specification]
hysteresis error 0 mm
Analog Output Accuracy (measured in accordance with IEC 60770-1)
non-linearity (accuracy) 0.100% of span (including hysteresis and repeatability)
non-repeatability 0.030% of span (included in non-linearity specification)
deadband (resolution) 0.030% of span [included in non-linearity specification]
hysteresis error 0%
Frequency 25 GHz nominal
Measurement range
1
0.35 to 100 m (1.15 to 328.08 ft).
Note: Siemens Milltronics makes every attempt to ensure the accuracy of these
specifications, but reserves the right to change them at any time.
1.
Referenced from flange face.